Summary
Type: SPE
Application #: 302280
Catalog#: 730012 - SPE Cartridges, Chromabond, C18 ec, 3ml, PP, 200mg, 50/pk
Detection: GC/MS
Conditions: Column conditioning: with 6 ml of methanol and 3 ml distilled water
Sample application: The two buffer fractions were passed trough the column.
Column washing: with 3 ml of water, followed by 3 ml of 0,6 M NaHCO3 and 3 ml of distilled water
Elution: The columns were dried and eluted with 2 ml of acetone - dichloromethane (3:1, v/v).
Evaporation: The eluant was evaporated to dryness under a stream of nitrogen at 40°C and reconstituted in 50μl of ethyl acetate.
Authors: Yegles, M.; Mersch, F.; Wennig, R.
Source: Forensic Science International 84 (1997) 211-218
Details
Legend: Our results in the hair of 21 subjects revealed the presence of nordazepam in 20 cases (95,2%), of diazepam in 15 cases (71,4%), of oxazepam in 15 cases (71,4%) and of aminoflunitrazepam in 8 cases (3